What it does
Converts torsional load on a transducer into a reading, for testing tools and rotating parts.
How it works
Taking a strain gauge transducer as an example, gauges on the twisted elastic element form a bridge, torque causes small deformation and resistance change, and the bridge signal is amplified and calibrated into torque. The test setup must also pass load along the specified torsional path so bending, axial force or fixturing changes do not enter the result.
Process sequence
- Check transducer and display match
- Fit interface and reaction support
- Confirm no load and zero
- Apply torque in the specified direction
- Read and save the result
- Unload and recheck zero
Key inputs
- Range and direction of torque
- Tool or specimen interface
- Static or dynamic loading
- Display match, calibration and extraneous loads
Common problems
- Loose or off-centre connections
- Overload causing permanent change
- Temperature and zero drift
- Transducer and amplifier calibration mismatch
Limits of application
- Transducer capability and test loading method together determine results
- Bridge self-checks cannot replace calibration
- Test joints and dynamic response must be confirmed separately for hand and power tools
Functional modules
Elastic element and strain bridge
Sense twist and output a signal
Amplification and display
Convert to a torque reading
Loading interface and reaction support
Transmit the specified load
Calibration link
Keeps the measuring chain matched
Manufacturer & technical sources
Continue with manufacturer product pages, operating principles and technical information for this equipment type.