What it does
Measures part response in tension, compression, torsion and similar processes under controlled linear or rotary loading.
How it works
A linear stand moves the grip relative to the force gauge, and a rotary stand turns the loading plate relative to the torque sensor. Grips keep load direction and specimen location, and readings can be matched to travel or angle; stand stiffness, gripping and loading method also affect readings.
Process sequence
- Fit suitable grips and sensor
- Centre and grip the specimen
- Zero and set loading conditions
- Load by displacement or angle
- Stop and unload at the specified condition
- Save the curve and judgment
Key inputs
- Measuring range and load direction
- Specimen gripping and allowable deformation
- Loading speed and end point
- Displacement or angle acquisition and system compliance
Common problems
- Off-centre loading adding extra moments
- Grip slippage
- Stand and sensor deflection affecting displacement
- Sampling and loading speed mismatch
Limits of application
- Force stands and torque stands differ in motion and sensors and are not interchangeable
- Gauge accuracy is not the test uncertainty of the whole setup
- Destructive tests need tooling that contains stored energy and fragments
Functional modules
Loading mechanism
Provides linear or rotary motion
Sensor and display
Measure force or torque
Grips and stand
Hold the specimen and load direction
Travel, angle and control
Organize end points and acquisition
Manufacturer & technical sources
Continue with manufacturer product pages, operating principles and technical information for this equipment type.